Accuracy of 2 Rapid Antigen Tests During 3 Phases of SARS-CoV-2 Variants

11 Nov.,2023

 

Key Points

Question  Are rapid antigen tests analytically and clinically accurate for detecting variants of SARS-CoV-2?

Findings  In this diagnostic study of 802 adults reporting COVID-19–like symptoms within the prior 5 days, no significant differences were found in the analytical limit of detection or clinical diagnostic accuracy of 2 rapid antigen tests across 3 epidemic phases of SARS-CoV-2 variants. The positive percent agreement ranged from 81% to 91% across the 3 phases of variants.

Meaning  This study found that 2 rapid antigen tests had consistent analytical and clinical accuracy across 3 phases of circulating SARS-CoV-2 variants.

Abstract

Importance  Variants of SARS-CoV-2 have sequence variations in the viral genome that may alter the accuracy of rapid diagnostic tests.

Objective  To assess the analytical and clinical accuracy of 2 rapid diagnostic tests for detecting SARS-CoV-2 during 3 phases of variants.

Design, Setting, and Participants  This diagnostic study included participants aged 18 years or older who reported onset of COVID-19–like symptoms within the prior 5 days and were tested at multiple COVID-19 testing locations in King County, Washington, from February 17, 2021, to January 11, 2022, during 3 distinct phases of SARS-CoV-2 infection (pre-Delta, Delta, and Omicron).

Interventions  Two anterior nasal swab specimens were collected from each participant—1 for onsite testing by the SCoV-2 Ag Detect Rapid Self-Test and 1 for reverse transcriptase–polymerase chain reaction (RT-PCR) testing.

Main Outcomes and Measures  The analytical limit of detection of the 2 rapid diagnostic tests (SCoV-2 Ag Detect Rapid Self-Test and BinaxNOW COVID-19 Ag Card) was assessed using Omicron (B.1.1.529/BA.1), Delta (B.1.617.2), and a wild-type (USA-WA1/2020) variant. Diagnostic sensitivity and specificity of clinical testing for the rapid antigen tests were compared with that of RT-PCR testing.

Results  A total of 802 participants were enrolled (mean [SD] age, 37.3 [13.3] years; 467 [58.2%] female), 424 (52.9%) of whom had not received COVID-19 vaccination and presented a median of 2 days (IQR, 1-3 days) from symptom onset. Overall, no significant differences were found in the analytical limit of detection or clinical diagnostic accuracy of rapid antigen testing across SARS-CoV-2 variants. The estimated limit of detection for both rapid nucleocapsid antigen tests was at or below a 50% tissue culture infectious dose of 62.5, and the positive percent agreement of the SCoV-2 Ag Detect Rapid Self-Test ranged from 81.2% (95% CI, 69.5%-89.9%) to 90.7% (95% CI, 77.9%-97.4%) across the 3 phases of variants. The diagnostic sensitivity increased for nasal swabs with a lower cycle threshold by RT-PCR, which correlates with a higher viral load.

Conclusions and Relevance  In this diagnostic study, analytical and clinical performance data demonstrated accuracy of 2 rapid antigen tests among adults with COVID-19 symptoms across 3 phases of SARS-CoV-2 variants. The findings suggest that home-based rapid antigen testing programs may be an important intervention to reduce global SARS-CoV-2 transmission.

Introduction

More than 500 million cases of confirmed SARS-CoV-2 infection and 6.2 million deaths from COVID-19 have been reported to the World Health Organization,1 and these numbers may be vastly underestimated.2,3 Implementation of diagnostic testing for acute SARS-CoV-2 infection has been essential to identify COVID-19 cases, reduce transmission, and inform public health measures.4,5 Initially, SARS-CoV-2 testing relied on laboratory-based reverse transcriptase–polymerase chain reaction (RT-PCR) of nasopharyngeal specimens.6 Although results of nucleic acid amplification tests have been critical, they have remained expensive and technically and logistically challenging.7,8

Rapid diagnostic tests can be used to diagnose acute SARS-CoV-2 infection.9-11 However, antigen-based assays that rely on antibody conjugation for capture and detection of viral proteins have not been universally endorsed for diagnostic testing.12-14 Rapid antigen–based diagnostic tests have now become widely abundant15 and may be useful to facilitate testing in community- and home-based settings, expedite treatment initiation, and optimize isolation periods.5,8,16-18

Variants of SARS-CoV-2 have sequence variations in the viral genome that may alter the accuracy of rapid diagnostic tests.5 Molecular tests can be affected by single-point sequence variations, whereas antigen tests may require multiple sequence variations to change the confirmation of viral protein epitopes. The Omicron variant has numerous sequence variations in the spike and nucleocapsid proteins,19 which has raised concerns from the US Food and Drug Administration about the analytical and clinical accuracy of rapid antigen–based diagnostic tests.20 Therefore, this study assessed the analytical and clinical accuracy of 2 rapid diagnostic tests for detecting SARS-CoV-2 variants (Delta and Omicron) involved with the predominant waves of the COVID-19 pandemic.

Methods

This diagnostic study assessed the analytical accuracy of 2 rapid antigen tests approved by the US Food and Drug Administration—SCoV-2 Ag Detect Rapid Self-Test (InBios International Inc) and BinaxNOW COVID-19 Ag Card (Abbott Laboratories)—using 3 replication-competent variants or strains, including Omicron (B.1.1.529/BA.1), Delta (B.1.617.2), and a wild type of SARS-CoV-2 (USA-WA1/2020). The study received ethical approval from the University of Washington, and participants provided verbal informed consent. The reporting of results followed the Standards for Reporting of Diagnostic Accuracy (STARD) reporting guideline.

The SARS-CoV-2 variants with a known 50% tissue culture infectious dose (TCID50) were obtained from BEI Resources. The reported TCID50 values for these variants were 4.4 × 105 TCID50/mL for Omicron (BA.1), 1.1 × 106 TCID50/mL for Delta (B.1.617.2), and 1.6 × 106 TCID50/mL for the USA-WA1/2020 strain. The sequences of these variants are available from the Global Initiative on Sharing Avian Influenza Data (BA.1 strain: EPI_ISL_7160424; B.1.617.2 strain: EPI_ISL_2103264) and/or the National Institutes of Health’s GenBank (USA-WA1/2020 strain: MN985325.1). The TCID50 value depicts the replication competence of the virus, which was further assessed by measuring the infectious viral particles as plaque-forming units through the infection of VERO E6-ACE2 and TMPRSS2 in a standard plaque assay.21

Performance and the analytical limit of detection for the Omicron variant were measured by spiking negative clinical nasal swab matrices with replication-competent virus to establish 3 stock viral concentrations of 5.0 × 104 TCID50/mL, 1.25 × 104 TCID50/mL, and 3.12 × 103 TCID50/mL. From stock concentrations, 20 μL of sample was transferred onto nasal swabs to generate high, medium, and low viral concentrations of 1000 TCID50, 250 TCID50, and 62.5 TCID50 per swab, respectively. Viral dilutions across variants were tested in triplicate for both the SCoV-2 Ag Detect Rapid Self-Test and BinaxNOW COVID-19 Ag Card. The results were measured based on the visual signal intensity among serial dilutions.

To complement assessment of the analytical accuracy of the SCoV-2 Ag Detect Rapid Self-Test, a clinical diagnostic accuracy study among 802 participants was conducted at multiple testing locations in King County, Washington, from February 17, 2021, to January 11, 2022, during 3 distinct phases of SARS-CoV-2 infection (pre-Delta [February 17 to April 29, 2021], Delta [September 2 to November 30, 2021], and Omicron [December 13, 2021, to January 11, 2022]). The testing periods were determined based on Washington State’s Department of Health data on the circulation of predominant variants.22 Participants were aged 18 years or older and reported onset of COVID-19–like symptoms within the prior 5 days. Two anterior nasal swab specimens were collected from each participant—1 for onsite testing by the SCoV-2 Ag Detect Rapid Self-Test and 1 for RT-PCR testing. Sex, gender, race, and ethnicity were asked by the research staff and self-reported by the participants.

Statistical Analysis

Basic descriptive analyses for diagnostic accuracy were conducted, and 95% CIs were computed using the exact binomial (Clopper-Pearson) method. A person was considered COVID-19 positive if they reported COVID-19–like symptoms and had a laboratory-based RT-PCR test positive for SARS-CoV-2. Data were analyzed with R, version 4.2.0 (R Project for Statistical Computing).

Results

In the clinical study, 802 participants were enrolled (mean [SD] age, 37.3 [13.3] years; 467 [58.2%] female), 424 (52.9%) of whom had not received COVID-19 vaccination. Five of these participants (3 from the Delta phase and 2 from the Omicron phase) were missing RT-PCR results and were excluded from the analyses. Participants were tested a median of 2 days (IQR, 1-3 days) from symptom onset (Table 1). The pre-Delta phase consisted mostly of Alpha, Epsilon, and Gamma variants in this study population. Overall, there were no apparent differences in participants among the 3 study phases, with the exception of increasing rates of vaccination and collection of nasal specimens by a research coordinator during the Delta phase. Across the study period, the median cycle threshold value by RT-PCR among participants positive for SARS-CoV-2 was 26.2 cycles (IQR, 22.9-29.5 cycles).

The diagnostic accuracy of the SCoV-2 Ag Detect Rapid Self-Test was consistent across SARS-CoV-2 variants (Table 2). The positive percent agreement ranged from 81.2% (95% CI, 69.5%-89.9%) to 90.7% (95% CI, 77.9%-97.4%). Of importance, the negative percent agreement remained high throughout the study period, with a cumulative diagnostic specificity of 99.8% (95% CI, 99.1%-100%). The assay had improved sensitivity for swab specimens that had a lower cycle threshold by RT-PCR testing, which correlates with a higher viral load (Table 3). Among participants with a cycle threshold of 30 or less, the positive percent agreement was 97.9% (95% CI, 94.1%-99.6%). In addition, there were no apparent differences in clinical test performance when stratified by vaccination status or days since symptom onset.

In the assessment of analytical accuracy, the visual signals of the SCoV-2 Ag Detect Rapid Self-Test were positively associated with viral concentration (Figure 1). The estimated limit of detection for both rapid nucleocapsid antigen tests was at or below 62.5 TCID50. In a similar assessment, the dilutions across variants for the BinaxNOW COVID-19 Ag Card were also positively associated with viral concentration, and the estimated limit of detection was consistent across variants. By subsequent plaque assay, which determines the amounts of infectious virus, the limit of detection (62.5 TCID50) corresponded to approximately 51 plaque-forming units (Figure 2). For both rapid antigen assays, there were no significant differences in the analytical limit of detection across SARS-CoV-2 variants.

Discussion

In this diagnostic study, analytical and clinical performance data demonstrated accuracy of 2 rapid antigen tests for detecting SARS-CoV-2 during 3 phases of variants. Other studies have similarly demonstrated good analytical sensitivity for rapid antigen tests to detect Omicron23,24 and similar limits of analytical detection between the Omicron variant and the USA-WA1/2020 strain.25 A field study showed that the BinaxNOW COVID-19 Ag Card had good clinical accuracy for Omicron.26 Overall, this study demonstrated both analytical and clinical accuracy of rapid antigen testing across several variants of concern, including Omicron and Delta.

Strengths and Limitations

This study has strengths. The detection of circulating SARS-CoV-2 variants was representative of the community prevalence over a 12-month period.

This study also has limitations. It used a 5-day testing window from the onset of symptoms, which is the approved period of use for these rapid diagnostic tests but may not reflect the full spectrum of clinical testing. Additional clinical and analytical comparative studies may be warranted for other rapid antigen tests. The positive and negative predictive values, which are based on the prevalence of disease, may not be generalizable to other populations. In addition, the Omicron strain evaluated was B.1.1.529/BA.1, and the analyses did not include more recent Omicron strains, such as BA.2 or BA.2.12.

Conclusions

In this diagnostic study, analytical and clinical performance data demonstrated the preserved accuracy of 2 rapid antigen tests across SARS-CoV-2 variants among adults with COVID-19 symptoms. Rapid antigen tests may correlate with the recovery of replication-competent SARS-CoV-227 and appeared to retain accuracy across variants. Although more clinical studies are needed, the ongoing home-based rapid antigen testing programs may be an important intervention to reduce global SARS-CoV-2 transmission.

Back to top

Article Information

Accepted for Publication: July 7, 2022.

Published: August 24, 2022. doi:10.1001/jamanetworkopen.2022.28143

Correction: This article was corrected on October 3, 2022, to fix Dr Hau’s academic degree in the byline.

Open Access: This is an open access article distributed under the terms of the CC-BY License. © 2022 Drain PK et al. JAMA Network Open.

Corresponding Author: Paul K. Drain, MD, MPH, University of Washington, 325 Ninth Ave, HMC# 359927, Seattle, WA 98104-2499 (pkdrain@uw.edu).

Author Contributions: Dr Drain had full access to all of the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

Concept and design: Drain, Dalmat.

Acquisition, analysis, or interpretation of data: All authors.

Drafting of the manuscript: Drain, Dalmat, Uppal, Verma.

Critical revision of the manuscript for important intellectual content: Drain, Bemer, Morton, Dalmat, Abdille, Thomas, Hau, Green, Gates-Hollingworth, AuCoin, Verma.

Statistical analysis: Bemer, Dalmat, Thomas, Uppal.

Obtained funding: Drain, AuCoin.

Administrative, technical, or material support: Drain, Morton, Abdille, Hau, Green, Gates-Hollingworth.

Supervision: Drain, Morton, Abdille, Thomas, Hau, Gates-Hollingworth, AuCoin, Verma.

Conflict of Interest Disclosures: Dr Drain reported receiving grants from InBios International during the conduct of the study; research funding, paid to his institution, from the National Institutes of Health (NIH), the US Centers for Disease Control and Prevention (CDC), the Bill and Melinda Gates Foundation, Gilead Sciences, and Abbott; and consulting fees from Gilead Sciences, ThermoFisher, Cepheid, InBios International, Abbott, PATH, LumiraDx, and Alveo Technologies. Dr Uppal reported receiving financial support from InBioS LLC during the conduct of the study and outside the submitted work. Dr Hau reported receiving grants from the University of Nevada, Reno during the conduct of the study. Dr Gates-Hollingworth reported receiving grants from the University of Nevada, Reno during the conduct of the study. Dr AuCoin reported receiving fees from InBios International for a service contract to evaluate performance of the InBios SCoV-2 Ag Detect Rapid Self-Test during the conduct of the study; being the chief executive officer of a DxDiscovery Inc university startup pursuing development of a rapid diagnostic test for SARS-CoV-2; receiving grants from InBios International and the AuCoin laboratory; and having licensed technology from InBios International and the AuCoin laboratory for diagnosis of melioidosis outside the submitted work. No other disclosures were reported.

Funding/Support: The study was funded by InBios International Inc (Dr Drain).

Role of the Funder/Sponsor: The funder had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication.

Additional Information: SARS-CoV-2 isolate USA-WA1/2020 was deposited by the CDC and obtained through BEI Resources, National Institute of Allergy and Infectious Diseases (NIAID), NIH: SARS-Related Coronavirus 2, NR-52281. The following reagents were obtained through BEI Resources, NIAID, NIH: SARS-Related Coronavirus 2, Isolate hCoV-19/USA/MD-HP05285/2021 (Lineage B.1.617.2; Delta Variant), and NR-55671, Isolate hCoV-19/USA/MD-HP20874/2021 (Lineage B.1.1.529; Omicron Variant), NR-56461, contributed by Andrew Pekosz, PhD (Johns Hopkins Bloomberg School of Public Health).