Isobutane Chemical Ionization Mass Spectrographic Examination of Explosives

24 Mar.,2023

 

The detection of explosive residues in debris is difficult because of the thermal instability of many explosives along with the high sensitivity requirements of the analyses. The Isobutane chemical ionization (CI) mass spectra of common civilian and military explosives were obtained under different instrumental parameters. The intent of the study was to determine the feasibility of applying CI to residue detection. The CI spectra of the explosives 1,3,5-trinitro-1,3,5-triazocydohexane, 1,3,5,7-tetraazocyclooctane, and pentaerythritol tetranitrate were shown to be particularly sensitive to the conditions of source temperature and reagent gas pressure. These parameters were adjusted to yield the least complex CI spectra for the explosives studied. The simplicity of the CI spectra obtained makes it a feasible technique for detecting explosive residues in the presence of extraneous materials found in the acetone extracts of debris material. Placement of the extract into the direct probe of the CI mass spectrometer eliminates the need for prior chromatographic treatment of the extract and would optimize the high sensitivity of the CI technique.